3.0 SETTING UP THE TCAT-2LV CONTROLLER
3.1 Connect a suitable load to the LVDC output
3.2 Insert a type T thermocouple feedback sensor into the blue socket. Ensure that the sensor tip is
securely attached to the load at an appropriate location. Call our Engineering Applications Department
(1-800-452-8510) for more information on sensor placement.
3.3 Connect the AC power cord to the AC input module and to any suitable AC outlet.
3.4 The auxiliary AC output socket is available for additional equipment and is activated by the
input module power switch.
3
POWER
SWITCH
AC INPUT MODULE
AUXILIARY AC
OUTPUT SOCKET
AC FUSE
REPLACE WITH
6.3A SLO BLO ONLY
CONTROLLED LOW VOLTAGE
DC OUTPUT
THERMOCOUPLE
SOCKET
FOR FEEDBACK
SENSOR
REAR PANEL CONNECTIONS
EB TCAT-2LV Rev 1 08/14/07